Description
Calculate the time constant TCRH (s) for an RS-80 radiosonde relative humidity sensor. The time constant is a function of pressure. The pressure pt (mb) is interpolated from a pressure series at time t. For a tropical atmosphere.
Formula
TCRH=18.3−0.02921pt+3.31∗10−5p2t−1.4∗10−8p3tSource
From Hal cole, via Editsonde.